2Wire AU2700 User Manual

OLYMPUS  
DIAGNOSTIC SYSTEMS GROUP  
AU2700 & AU5400  
ON-LINE SPECIFICATIONS  
- 1 -  
Table of contents  
page  
1. Outline..............................................................................................................................................................................................4  
2. Basic Specification...........................................................................................................................................................................5  
2.1  
2.2  
2.3  
Format of Transmission....................................................................................................................................................5  
Transmission code............................................................................................................................................................5  
Text format.......................................................................................................................................................................6  
3.Transmission Text Format.................................................................................................................................................................9  
3.1  
3.2  
3.3  
3.4  
3.5  
3.6  
Common and changeable item..........................................................................................................................................9  
T.R.I. inquiry text .............................................................................................................................................................9  
T.R.I. text .......................................................................................................................................................................11  
Test result transmission text ...........................................................................................................................................12  
Relation between real-time/batch and Transmission text................................................................................................21  
Others .............................................................................................................................................................................22  
4. Transmission Protocols ..................................................................................................................................................................30  
4.1  
4.2  
4.3  
Protocols at AU2700/5400 .............................................................................................................................................30  
Timing/time-out..............................................................................................................................................................33  
Rule of class B................................................................................................................................................................35  
5.Application Protocol .......................................................................................................................................................................37  
5.1  
5.2  
5.3  
Receiving T.R.I. .............................................................................................................................................................37  
Sending test result protocol ............................................................................................................................................40  
Others .............................................................................................................................................................................42  
6. Specification of Connection ...........................................................................................................................................................43  
6.1  
6.2  
6.3  
6.4  
In/Output signals and Terminals.....................................................................................................................................43  
Signal Level....................................................................................................................................................................43  
Applicable Connector.....................................................................................................................................................43  
others ..............................................................................................................................................................................43  
A. Appendix…………………………………………………………………………………………………………………..44  
1. Flags list ...............................................................................................................................................................................44  
2. Online parameters list...........................................................................................................................................................46  
3. Character table......................................................................................................................................................................49  
4. Online parameter sheet ........................................................................................................................................................50  
5. Online alarm list ...................................................................................................................................................................52  
- 3 -  
1. Outline  
1) The following data is exchanged between the AU2700/5400 and the host computer system.  
(1) Test Requisition Information (T.R.I.)  
(2) Test Results  
2) The data transmission is conducted in either of the following mode:  
(1) Class A Transmission is without synchronizing with analysis. This mode does not utilize the ACK/NAK protocol.  
(2) Class B Transmission is synchronizing with analysis. This mode does utilize the ACK/NAK protocol.  
3) The Following format and protocol can be selected by [PARAMETER] [ONLINE] menu.  
(1) Setup  
(2) Communication protocol  
A. T.R.I. Inquiry Text Format  
B. Test Result Text Format  
(3) Test Number Assignment  
2. Basic Specification  
- 4 -  
(2.1) Format of Transmission  
Item  
Contents  
RS-232C  
Transmission  
line  
Synchronization  
Synchronization  
Half-Duplex  
Data transmission  
mode  
Bit/sec  
4800bps, 9600bps  
Configuration  
start bit  
data bit  
parity bit  
stop bit  
total  
1 bit  
7 or 8 bit  
none, odd or even  
1 or 2 bit  
9-12 bit  
These can be selected by [PARAMETER] [ONLINE] [PROTOCOL]  
menu.  
Form of  
Transmission  
Class A Transmission are conducted from the sender to  
the receiver in a certain time of interval.  
Class B Transmission are conducted by confirming with ACK  
(receipt acknowledged)or NAK (not acknowledged)  
between the sender and the receiver.  
1 channel  
Use channel  
Retry  
Class A: None  
Class B: 0-3 This can be selected by [PARAMETER] [ONLINE]  
[PROTOCOL]menu.  
(2.2) Transmission code  
Item  
Contents  
Limit of value  
20H-7EH  
Data code  
7 bit code ASCII CODE  
8 bit code  
These can be  
selected by  
[PARAMETER]  
[ONLINE]  
[PROTOCOL]  
menu.  
1 byte code  
20H-7EH  
A1H-DFH  
Control code  
Start code/terminate code  
01H-1FH  
06H  
15H  
ACK  
NAK  
BCC  
00H-FFH  
- 5 -  
(2.3) Text format  
(1) Basic configuration as follows:  
1) 2) 3) 4)  
5) 6)  
Name  
Digit  
Contents  
Remarks  
1) Text start code  
2) Text  
1,2  
2
01h-1Fh  
R#  
Data start code  
Normally 02h  
AU2700/5400  
Host  
T.R.I. inquiry  
Classification  
RB  
R_ *  
Start of T.R.I. inquiry  
T.R.I. inquiry of  
normal sample  
RH  
T.R.I. inquiry of  
rerun sample  
RE  
End of T.R.I. inquiry  
T.R.I.  
S#  
Host  
S_ *  
SH  
SE  
T.R.I. of normal sample  
T.R.I. of rerun sample  
Stop of T.R.I. inquiry  
Test result  
AU2700/5400  
D#  
AU2700/5400  
DB  
Start of transfer result  
Normal sample result  
Rerun sample result  
Reagent blank sample  
Result  
Host  
D_ *  
DH  
DR  
DA  
Calibration result  
d_ *  
Stat fast result(AU2700 Only)  
Emergency fast result  
(AU5400 Only)  
dH  
Stat fast rerun result  
(AU2700 Only)  
Emergency fast rerun result  
(AU5400 Only)  
DQ  
DE  
QC sample result  
End of result transfer  
3) Unit No.  
0,2  
[PARAMETER]  
[ONLINE] [PROTOCOL]  
menu.  
(Optional)  
4) Text code  
Contents of Text  
Data No. is added behind  
text of Header parts.  
[PARAMETER]  
[ONLINE] [PROTOCOL]  
menu.  
5) text end code  
1,2  
1
01h-1Fh  
00h-FFh  
End of text code  
[PARAMETER]  
[ONLINE] [PROTOCOL]  
menu.  
6) BCC  
The sum with the Exclusive  
or logic between 2-5 and including  
5.  
[PARAMETER]  
[ONLINE] [PROTOCOL]  
menu.  
(Block Check  
Character)  
(Optional)  
remarks * _ is space.  
- 6 -  
(2) Blocking  
A. Definition of Term  
1) 2) 3)  
4)  
5) 6)  
Term  
Definition  
Is the total number of bytes from 1) to 6).  
Text length  
[PARAMETER]  
[ONLINE]  
[PROTOCOL] menu.  
Max block  
length  
It is the Max length of text in one record.  
The length of block can be selected in online parameter menu.  
(256, 512, 1024)  
[PARAMETER]  
[ONLINE]  
[PROTOCOL] menu.  
- 7 -  
B. Kind of text and blocking  
Kind of text  
Kind  
Method of Blocking  
Non-blocking  
Fixed length  
Text  
Start of T.R.I. inquiry  
T.R.I. inquiry text  
End of T.R.I.  
transmission text.  
Stop of T.R.I. inquiry  
Start of transfer result  
transmission text.  
End of result transfer  
transmission text.  
Variable  
length text  
Using  
blocking  
Max block length  
< text length  
T.R.I. text  
Rerun T.R.I. text  
Test result text  
Block  
Identification  
No.  
first block  
Block  
Identification No. =0  
second block  
Rerun result text  
QC result text  
Acal result text  
Reagent blank result  
text  
(0-9,E)  
Block  
Identification No. =1  
STAT(fast) result text  
(AU2700 only)  
Emergency(fast) result  
text  
(see page 9)  
(AU5400 only)  
last block  
Block  
Identification No. =E  
Text end  
code  
Text end code = ETX(03H)  
and using ETB = yes *1  
End of block = ETB  
End of text = ETX  
Text end code <>ETX(03H)  
or ETB(17h)is no-use *1  
End of block is  
designated in menu.  
End of text is  
designated in menu.  
Using ETB can be selected  
in online parameters.  
Remark *1:When Block Identification No. is numeric(0,1,....), Only “End of Block” is used.  
When Block Identification No. is “E”, Only “End of Text” is used.  
- 8 -  
3. Transmission Text Format  
(3.1) Common and changeable item  
The following can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu.  
Text start code 1)  
Unit No. 3)  
Common Item  
Text end code  
BCC  
5)  
6)  
Sample Type  
Dummy(in Test Result transmission Text)  
Block Identification No.  
Rack No./Cup Position  
Unit No. on the T.R.I. text (with out Rerun)  
ANL unit No. and Cuvette side on Test Result transmission Text (without RB,ACAL,QC)  
Space Code before the Text Classification (S,R,D text)  
Reagent Lot No. and bottle No.  
The order of the Sample number and Sample ID on the text can be changed, the sample type, the Dummy spaces and  
the Block Identification Number can be changed in the [PARAMETER] [ONLINE] [PROTOCOL] menu. There is a  
function at the bottom of the page called “Special”. These features may be altered under the “Special” menu.  
Making changes in the “Special” menu only affects the T.R.I. Text and Test Result Transmission Text.  
(3.2) T.R.I. inquiry text  
(1) Start text to inquiry test requisition.  
1) R B 3) 5) 6)  
(2) Normal run inquiry test requisition.  
Cup  
1) R  
_
3) Rack no.  
Posi-  
tion  
Sample no.  
Sample ID  
5) 6)  
Sample type  
- 9 -  
(3) Repeat run inquiry test requisition.  
Cup  
1) R H 3) Rack No.  
Posi-  
tion  
Sample no.  
Sample ID  
5) 6)  
Sample type  
(4) End text to inquiry test requisition.  
1) R E 3) 5) 6)  
- 10 -  
(3.3) T.R.I. text  
(1) Normal sample test requisition information text.  
cup  
1) S _  
Rack No.  
posi-  
tion  
Sample no.  
Sample ID  
Dummy  
_
_
_
_
Sample type  
Block Identification No.  
s
Patient  
Patient  
Patient  
e
age  
month  
Information 1  
Information 2  
Information 3  
x
Patient  
Information 4  
Patient  
Information 5  
Patient  
Information 6  
Test Test Test Test Test Test Test  
code code code code code code code  
Test  
code 5) 6)  
(2) Repeat run sample test requisition information text.  
Cup  
Original  
1) S H Rack no.  
Posi-  
tion  
Sample no.  
Sample ID  
Sample no.  
Sample type  
Block Identification No.  
Blocking  
Test Me Test Me Test Me Test Me  
code th code th code th code th  
Test Me  
code th 5) 6)  
od  
od  
od  
od  
od  
- 11 -  
(3) End of T.R.I. Text transmission.  
1) S E 5) 6)  
(3.4) Test result transmission text  
(1) Start of test result transmission text.  
1) D B 3) 5) 6)  
- 12 -  
(2) Normal sample test result text.  
Cup  
1) D  
_
3) Rack no.  
Posi-  
tion  
Sample no.  
Sample ID  
Dummy  
_
_
_
_
Sample type  
Block Identification No.  
S
e
Patient  
Information 1  
Patient  
Information 2  
Patient  
Information 3  
Age  
Month  
x
Patient  
Patient  
Patient  
Information 4  
Information 5  
Information 6  
Blocking  
Test  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
code  
Cuvette side  
ANL unit No.  
Blocking  
Flags  
 Test Result  
5) 6)  
- 13 -  
(3) Stat(fast) test result text (AU2700 only).  
Cup  
P n n n  
1)  
d
_ 3) Rack no.  
Posi-  
tion  
Sample no.  
Sample ID  
Dummy  
_
_
_
_
_
_
_
_
1-22  
Sample type  
Block Identification No.  
S
e
Patient  
Information 1  
Patient  
Information 2  
Patient  
Information 3  
Age  
Month  
x
Patient  
Patient  
Patient  
Information 4  
Information 5  
Information 6  
Blocking  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
Test  
code  
Cuvette side  
ANL unit No.  
Blocking  
Flags  
 Test Result  
5) 6)  
- 14 -  
(4) Emergency(fast) test result text (AU2700 only).  
Cup  
Posi-  
tion  
E n n n  
Sample no.  
1) d _ 3) Rack no.  
Sample ID  
Dummy  
_
_
_
_
_
_
_
_
1-22  
Sample type  
Block Identification No.  
S
e
Patient  
Information 1  
Patient  
Information 2  
Patient  
Information 3  
Age  
Month  
x
Patient  
Patient  
Patient  
Information 4  
Information 5  
Information 6  
Blocking  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
Test  
code  
Cuvette side  
ANL unit No.  
Blocking  
Flags  
 Test result  
5) 6)  
- 15 -  
(5) Repeat run result text -1 .  
1) D H 3) Rack no.  
Cup  
Posi-  
tion  
Original  
Sample no.  
Sample no.  
Sample ID  
Sample type  
Block Identification No.  
Blocking  
Test  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
code  
Cuvette side  
ANL unit No.  
Blocking  
Flags  
 Test result  
5) 6)  
- 16 -  
(6) Repeat run result text -2 Stat fast sample (AU2700 Only) .  
Cup  
Posi-  
tion  
P n n n  
Sample no.  
Original  
Sample No.  
1) d H 3) Rack no.  
_ _ _ _  
Sample ID  
1-22  
Sample type  
Block Identification No.  
Blocking  
Test  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
code  
Cuvette side  
ANL unit No.  
Blocking  
Flags  
 Test result  
5) 6)  
- 17 -  
(7) Reagent blank result text.  
Dummy  
_ _  
1) D R 3)  
_
_
_
_
_ _  
Sample no.  
_
_
_
_
_
_
_
Sample type  
This dummy is same ID digits  
and assigned area is decided associated  
with the rack No.  
Block Identification No.  
ANL unit No.  
Cuvette side  
Test  
code  
R1 Lot No.  
R1 Bottle No.  
R2 Lot No.  
R2 Bottle No.  
1:First data  
2:Second data  
Test Result  
Flags  
5) 6)  
- 18 -  
(8) Acal (Calibration) result text.  
Dum  
my  
1) D A 3)  
_
_
_
_
_ _  
Sample no.  
Calibrator ID  
_ _  
This dummy spaces is assigned in case rack No. exists  
and assigned area is decided associated with the rack No.  
Calibrator No. (1-80)  
Block Identification No.  
ANL unit No.  
Cuvette side  
Blocking  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
Test  
Code  
Blocking  
Flags  
 Test result  
5) 6)  
- 19 -  
(9) QC (Quality Control) result text.  
Dum  
my  
1) D Q 3)  
_
_
_
_
_
_
Sample no.  
Control ID  
_
_
This dummy No. is assigned in case rack No. exists  
Control No.  
and assigned area is decided associated with the rack No.  
Block Identification No.  
ANL unit No.  
Cuvette side  
Blocking  
R1 Bottle  
R1 Lot No. No.  
R2 Bottle  
R2 Lot No. No.  
Test  
code  
Blocking  
Flags  
 Test result  
5) 6)  
(10)End of test result transfer text.  
1) D E 3) 5) 6)  
- 20 -  
(3.5) Relation between real-time/batch and Transmission text  
T.R.I. inquiry of Routine, Emergency, Stat and Repeat run can be selected as real-time or batch.  
Test results of all samples can be selected as real-time or batch.  
- 21 -  
(3.6) Others  
(1) Contents and formats of text.  
Item  
Digit  
Contents  
Remarks  
1. Rack No.  
4 or 5  
'0001'-'9999' or  
'00001'-'99999'  
STAT sample is space  
The digit of rack no.  
and cup pos. and  
these using is  
changeable in  
[PARAMETER]  
[ONLINE]  
[PROTOCOL]  
menu.  
(Optional)  
When Rack no. is no setting,  
Rack No. is '00000'.  
In case ,receive  
T.R.I.  
in batch.  
2. Cup position  
(Optional)  
2
1
4
rack : '01'-'10'  
STAT : '01'-'22'  
This is automatically  
assigned when rack  
No. is in use.  
3. Sample type  
4. Sample No.  
space : serum  
U : urine  
X : other  
'0001'-'9999' : Routine sample  
'E001'-'E999' : Emergency sample  
'P001'-'P999' : Stat sample  
'R001'-'R999' : Reagent blank  
'A001'-'A999' : Calibration sample  
'Q001'-'Q999' : QC sample  
5. Sample ID  
4-26  
Number or character  
Number of digits can  
be changed in the  
[PARAMETER]  
[FORMAT]  
[FORMAT  
REQUISITION]  
menu.  
6. Original  
Sample No.  
4
'0001'-'9999' : normal sample  
'E001'-'E999' : emergency sample  
'P001'-'P999' : stat sample  
For Repeat Run Results.  
7. Reserved  
4
1
Space  
8. Block  
Identification  
No.  
'0'-'9' Last text is 'E'  
- 22 -  
9.ANL unit No.  
1
1
1
‘1’: AU5400 ISE  
‘1’: AU5400 unit 1  
‘2’: AU5400 unit 2  
‘3’: AU5400 unit 3  
ANL unit No. is added to the  
header portion of RB, ACAL,  
and QC result data. Normal  
sample and rerun sample, it is  
added to each item. However,  
in the case of an calculate  
item, a space is added. In a  
normal sample and repeat run,  
it can be prevented from  
adding a unit number by the  
setup of an on-line condition  
menu. This option can be  
selected in the  
(Optional)  
‘1’: AU2700  
[PARAMETER][ONLINE]  
[PROTOCOL][SPECIAL]  
menu.  
10.Cuvette side  
AU5400  
Cuvette side is added to the  
header portion of RB, ACAL,  
and QC result data. Normal  
sample and rerun sample, it is  
added to each item. However,  
in the case of an calculate  
item, a space is added. In a  
normal sample and rerun, it  
can be prevented from adding  
a unit number by the setup of  
an on-line condition menu.  
This option can be selected in  
the  
‘0’ : inner cuvette (ISE Cell 1)  
‘1’ : outer cuvette (ISE Cell 2)  
AU2700  
‘0’ : inner cuvette(ISE)  
‘1’ : outer cuvette  
(Optional)  
[PARAMETER][ONLINE]  
[PROTOCOL][SPECIAL]  
menu.  
11. Sex  
M : male  
These items can be  
selected/canceled in  
the [PARAMETER]  
[FORMAT]  
F : female  
SP : none sex  
0 : no-set  
(Optional)  
[FORMAT  
REQUISITION]  
menu.  
12. Age  
3
'000'-'150'  
space : no-set  
*1  
*1  
(Optional)  
13. Month  
(Optional)  
14. Patient  
Information  
(Optional)  
15. Online  
Test No.  
2
'00'-'11'  
space : no-set  
20  
number or character  
2
1
'01'-'99'  
This number can be  
changed in the  
[PARAMETER]  
[ONLINE] [TEST  
NO.]  
16. Method  
'0':normal  
For Repeat Run Results.  
'1':dilution  
‘2’:condense  
- 23 -  
17. Test result  
8 or 11  
Test Result digits can be 6 or 9.  
Flags digits is 2.  
0 suppress can be  
selected/canceled in  
online parameter  
menu  
Number of digits can  
be changed to  
6 or 9 in  
Refer to Appendix 1 (Flags List)  
online parameter  
menu.  
18. Lot No. and  
Bottle No.  
4*4  
Number or character  
This option can be  
selected in the  
(Optional)  
[PARAMETER]  
[ONLINE]  
[PROTOCOL]  
[SPECIAL] menu.  
Remark *1: If only age is the space in the text from a host, AU2700/AU5400 will output an error. The combination of Age  
and Month is as follows.  
Host -> AU2700/5400  
Pattern  
Age  
Month  
Spaces  
Spaces  
XX  
Result  
OK  
OK  
Error  
OK  
1
2
3
4
Spaces  
XXX  
Spaces  
XXX  
XX  
- 24 -  
(2) Test result format.  
A. Test result  
Data digit is 6 and 0 suppress is not used.  
0
1 2  
3
.
4
Data digit is 9 and 0 suppress is not used.  
-
-
-
0
_
_
1
2
3
.
4 5  
6
Data digit is 6 and 0 suppress is used.  
1
2 3  
4
Data digit is 9 and 0 suppress is used.  
1
2 3  
.
4
5
6
- 25 -  
B. LIH result  
Data digit is 6 and 0 suppress is not used.  
ICT HEM  
LIP  
0
0
0
0 NORMAL  
1 +  
FLAG  
2 ++  
3 +++  
4 ++++  
5 +++++  
6 ABN  
9 Test not performed.  
Data digit is 9 and 0 suppress is not used.  
LIP  
ICT  
HEM  
0 0  
0 0  
0 0  
0 NORMAL  
1 +  
FLAG  
2 ++  
3 +++  
4 ++++  
5 +++++  
6 ABN  
9 Test not performed.  
- 26 -  
Data digit is 6 and 0 suppress is used.  
ICT HEM  
LIP  
_
_
_
0 NORMAL  
1 +  
FLAG  
2 ++  
3 +++  
4 ++++  
5 +++++  
6 ABN  
9 Test not performed.  
Data digit is 6 and 0 suppress is used.  
LIP  
ICT  
HEM  
_ _  
_ _  
_ _  
0 NORMAL  
1 +  
FLAG  
2 ++  
3 +++  
4 ++++  
5 +++++  
6 ABN  
9 Test not performed.  
- 27 -  
(3) T.R.I. of calculated tests.  
Note: Calculated Test No. in T.R.I. from Host is ignored. The tests that create the calculated tests  
can only be ordered from the Host.  
(4) Data format that Test result data is digit over.  
a. Data digit is 6.  
9
9
9
9
9 9  
Data digit is 9.  
9
9
9
9
9
9
9
9
9
(5) Data format that Test result data is OD-value.  
Data digit is 6.  
0
.
1 2  
3
4
When the data is negative OD-value, it is same as digit over.  
Data digit is 9.  
-
0
.
1
2
3
4
- 28 -  
(6) T.R.I of LIH.  
In any case, LIH test no. in T.R.I from Host is necessary for LIH analysis.  
LIH Reagent Test Name  
LIH  
LIH Measure  
All  
Even if AU2700/AU5400 does not have the request from  
a host, it performs LIH analysis.  
Select  
All  
AU2700/AU5400 performs LIH analysis according to the  
request from a host.  
Even if AU2700/AU5400 does not have the request from  
a host, it performs LIH analysis.  
Other  
No  
LIH analysis is not performed even if AU2700/AU5400  
has the request from a host.  
Note 1) These can be selected by the following menus.  
LIH Reagent Test Name : [PARAMETER] – [COMMON TEST PARAMETERS] – [TEST NAME]  
LIH Measure  
: [PARAMETER] – [COMMON TEST PARAMETERS] – [ROUND]  
- 29 -  
4. Transmission Protocols  
(4.1) Protocols at AU2700/5400  
(1) Class A: without ACK/NAK exchange.  
A. Sending  
Text to be sent:  
T.R.I. Inquiry text  
Repeat Run T.R.I. Inquiry text  
Start of Test Result Transmission text  
Test Result text  
Repeat Run Result text  
End of Test Result Transmission text  
Start  
of  
Text  
Block end (BCC)  
Text end (BCC)  
Initial  
Condition  
Transmission text  
B. Receiving  
Text to be received:  
T.R.I. text  
Repeat Run T.R.I. text  
End of T.R.I. Transmission text  
Start  
of  
Text  
Block end (BCC)  
Text end (BCC)  
Initial  
Condition  
Transmission text  
- 30 -  
(2) Class B: with ACK/NAK protocol.  
A. Sending  
Text to be sent:  
T.R.I. Inquiry text  
Repeat Run T.R.I. Inquiry text  
Start of Test Result Transmission text  
Test Result text  
Repeat Run Result text  
End of Test Result Transmission text  
ACK  
ACK  
Received  
Received  
Block end (BCC)  
Initial  
Condition  
Start  
of  
Transmission text  
Text  
Text end (BCC)  
NAK  
Received  
or  
NAK  
Received  
or  
1st time: resending the same block  
2nd time: sending the next block  
Time-out  
Time-out  
1st time: resending the same text  
2nd time: return to initial condition  
    
- 31 -  
B. Receiving  
Test to be sent:  
T.R.I. text  
Repeat Run T.R.I. text  
End of T.R.I. Transmission text  
ACK  
Send  
ACK  
Send  
Block end (BCC)  
Initial  
Condition  
Start  
of  
Transmission text  
Text  
Text end (BCC)  
NAK  
Send  
NAK  
Send  
1st time: receiving the same block  
2nd time: receiving the next block  
1st time: receiving the same text  
2nd time: return to initial condition  
- 32 -  
(4.2) Timing/time-out  
(1) Time-out/timing list .  
Kind  
T1  
Description  
Default  
2 s  
Remarks  
Limit time from end of transfer/receive to start  
of text receive.  
Limit time from start of text receive to end of  
text receive.  
Limit time from start of text transfer to end of  
text transfer.  
Limit time from end of text transfer to response of  
receive.  
Interval time of transmission  
Min. time from receive  
Limit time from NAK response to start receive  
of retry text.  
0.1 x n sec  
(Note 2)  
T2  
T3  
T4  
(Note 1)  
n=1-99  
2 s  
2 s  
1 s  
2 s  
T5  
T6  
T7  
t1  
t2  
Min. time from end of transfer text to receivable  
text.  
Min. time from end of transfer text to next receivable  
text.  
Min. time from end of transfer text to receivable  
Response.  
Min. time from end of receiving text to transfer  
Response.  
0.5 s  
Not Changeable  
0.5 s  
0.5 s  
0.5 s  
t3  
t4  
note 1) ((Max text length x Character length)/Bit/sec)+ 0.5 s  
note 2) These time-outs can be changed in the online parameters menu.  
(2) Rule of class A  
A. Case 1  
A
C
AU2700/5400  
Host  
B
<T3  
A
<T2  
B
<T3  
C
t1<<T1  
T5<  
B. Case 2  
AU2700/5400  
Host  
<T3  
<T3  
<T3  
T5<  
T5<  
- 33 -  
C. Case 3  
AU2700/5400  
Host  
A
B
C
<T2  
<T2  
<T2  
t2<<T1  
t2<<T1  
- 34 -  
(4.3) Rule of class B  
(1) Case 1  
A
A
C
K
C
AU2700/5400  
Host  
A
C
K
B
<T3  
<T2  
<T3  
t3<<T4 t1<<T1  
t4<  
T5<  
(2) Case 2  
A
B
AU2700/5400  
Host  
A
C
K
<T3  
<T3  
t3<<T4  
T5<  
(3) Case 3  
A
C
K
AU2700/5400  
Host  
A
C
<T2  
<T2  
t4<  
t2<<T1  
(4) Case 4 (Abnormal Case 1)  
A
A
B
AU2700/5400  
Host  
N
A
K
N
A
K
<T3  
<T3  
<T3  
t3<<T4  
T6<  
t3<<T4  
T5<  
- 35 -  
(5) Case 5 (Abnormal Case 2)  
A
A
AU2700/5400  
Host  
N
A
K
N
A
K
B
<T3  
<T3  
<T2  
t3<<T4  
T6<  
t3<<T4  
t1<  
(6) Case 6 (Abnormal Case 3)  
N
A
K
N
A
K
AU2700/5400  
Host  
A
A
B
<T2  
<T2  
<T2  
t4<  
t2<<T7  
t4<  
t1<<T1  
(7) Case 7 (Abnormal Case 4)  
N
A
K
N
A
K
B
AU2700/5400  
Host  
A
A
<T2  
<T2  
<T3  
t4<  
t2<<T7  
t4<  
T5<  
- 36 -  
5. Application Protocol  
(5.1) Receiving T.R.I.  
(1) Trans./receive sequence in one session.  
General sequence :  
A. Example 1  
Shift next Rx-transfer when Online Error occurred.  
RB  
Rx  
Rx  
Rx  
AU2700/5400  
Host  
・・・  
0
E
0
E
Sx  
Sx  
Sx  
Sx  
SE  
In case 2 blocks  
B. Example 2  
Stop the session when Online Error occurred.  
RB Rx Rx  
RE  
AU2700/5400  
Host  
・・・  
0
E
0
E
Sx  
Sx  
Sx  
Sx  
(2) Detail  
T.R.I.  
Receive Kind of text  
Timing/condition  
Real-time RB(request  
Start)  
This text is transferred when it is shifted from stand-by  
mode to start measure and then the first Rx is transmitted.  
R_ * (request  
normal)  
(Note 2)  
Request of  
sample no.  
Request of  
sample id  
RH (request  
rerun)  
When detecting cup, this Sample No. is transferred if this  
sample has not T.R.I..  
When reading ID is normally, this Sample ID is transferred  
if this sample has not T.R.I..  
(Note 2)  
Request of  
sample no.  
Request of  
sample ID  
When detecting cup, this Sample No. is always transmitted  
When reading ID is normally, this Sample ID is always  
transmitted.  
remarks * _ is space.  
Note 1) The number of times a T.R.I. inquiry can be sent when an online error is generated can be selected in  
[PARAMETER] [ONLINE] [PROTOCOL].  
Note 2) It is possible that the R_-text and RH-text are mixed in one session.  
- 37 -  
Receive  
Kind of text  
Timing/condition  
This text can be received after end of Rx-transfer within  
the designated time.  
Real-time Sx  
SE  
RE (request  
This text is transferred when shifting to the modes below.  
1) measure mode to standby mode  
end)  
2) measure mode to stop mode  
This text is transmitted when the communication is  
canceled by the online alarm.  
Batch  
RB (request  
start)  
This text is transmitted when receiving T.R.I. is sent in the test  
requisition.  
R_ * (request  
Normal)  
(Note 3)  
Request of  
sample no.  
This text is transmitted at the designated time intervals.  
The texts are the samples that are selected in the test  
requisition menu.  
RH (request rerun)  
(Note 3)  
Request of  
sample no.  
This text is transmitted at the designated time intervals.  
The texts are the samples that are selected in the test  
requisition menu.  
Sx  
This text can be received after end of Rx-transfer within  
the designated time.  
SE  
RE (request end)  
This text is transmitted when it was receiving the last sample No.  
in test requisition menu and SE text isn't received.  
This text is transmitted when the communication is  
canceled in the test requisition menu.  
This text is transmitted when the communication is  
canceled by the online alarm.  
remarks * _ is space.  
Note 3) R_-text and RH-text is transferred in separate sessions.  
- 38 -  
(3) Kind of T.R.I. transfer/receive.  
A. T.R.I. of normal sample  
Setting of parameters  
Method of T.R.I. inquiry & T.R.I.  
Receive T.R.I.. inquiry  
Method of  
Kind of T.R.I.  
T.R.I. inquiry  
T.R.I. inquiry  
Sample no  
T.R.I.  
Real-time  
Sequential  
Rack no.  
Request  
of sample no  
Sample no *1  
Sample no  
(Calculated rack  
No. and cup pos.  
No.)  
Sample ID  
Request  
of sample ID  
Request  
Sample ID  
Sample no *1  
Sample no.  
Sample ID *1  
Sample no *1  
Sample no *1  
Batch  
Sequential  
Rack no.  
of sample no  
Sample ID  
Sample ID *1  
Sample no *1  
Selectable setting of receiving T.R.I. in online parameter menu and requisition format menu.  
Selectable method of T.R.I. in system menu.  
Remark *1: Sample no. and Sample ID of T.R.I. must be same to the Sample no. and Sample ID of T.R.I inquiry.  
(4) T.R.I. of Repeat sample  
Setting of parameters  
Repeat sample  
Receive of T.R.I. Repeat rack  
Method of T.R.I. inquiry & T.R.I.  
Method of  
T.R.I. inquiry  
Kind of T.R.I.  
T.R.I. inquiry  
T.R.I.  
Real-time  
Using  
Sequential  
rack no.  
Request  
of Sample no.  
Original  
sample no. *1  
Repeat sample no.  
Repeat sample no.*1  
Sample ID *1  
Sample ID  
Request  
Sample ID  
of sample ID  
Repeat sample no.  
*1  
Repeat sample no.*1  
Batch  
-
Sequential  
rack no.  
Request  
of sample no.  
Repeat sample no.  
Original  
sample no. *1  
Sample ID  
Sample ID *1  
Remark *1: Repeat sample no. and sample ID of T.R.I. must be same as the Repeat sample no. and sample ID of T.R.I  
inquiry.  
- 39 -  
(5.2) Sending test result protocol  
(1) Transfer sequence for one session.  
General sequence :  
A. Example 1  
Shift next Dx-transfer when Online Error occurred.  
DB Dx Dx Dx Dx  
Dx  
DE  
AU2700/5400  
Host  
0
E
0
E
E
2 blocks  
B. Example 2  
Stop session when Online Error occurred.  
DB  
Dx  
Dx  
Dx  
Dx  
Dx  
DE  
AU2700/5400  
Host  
0
E
0
E
E
(2) Detail  
Data  
Trans. Kind of text  
Timing/condition  
Normal end  
Online error  
Real  
-time  
DB (START)  
This text is transmitted when  
it is shifted from stand-by  
mode to measure mode and just  
It is shifting the  
transfer of the  
test results.  
It is output the  
alarm and it is  
executed the below  
behind the first Dx is transmitted.  
Error protocol is  
stop.  
It is canceled the  
transfer of Dx/DE  
Error protocol is  
continue  
Dx (result data) It is transmitted one by one  
the samples that are finished  
measuring.  
It is shifting next  
The transfer of  
Dx/DE.  
It is shifting next  
the transfer of  
Dx/DE.  
DE (end)  
This text is transmitted after  
the last transfer of Dx text  
and the below shifting mode.  
It is ended the  
session of transfer of alarm and ended  
The text of test  
results.  
It is output the  
the session of the  
transfer of the  
1) measure mode to standby  
2) measure mode to stop  
This text is transmitted when  
the communication is canceled  
by the online alarm.  
text of test results.  
- 40 -  
Data  
Trans.  
Kind of text  
DB (start)  
Timing/condition  
Normal end  
Online error  
Batch  
This text is transmitted when it  
selected the start of transmission  
in the online menu.  
It is shifting the  
transfer of the test  
Results.  
It is output the  
alarm and it is  
executed the below  
Error protocol is  
stop.  
It is canceled the  
transfer of Dx/DE  
Error protocol is  
continue.  
This text is transmitted at  
the designate time intervals.  
The text are the samples that  
is selected in the online menu  
It is shifting next  
The transfer of  
Dx/DE.  
Dx  
(result data)  
It is shifting next  
the transfer of  
Dx/DE.  
DE (end)  
This text is transmitted when  
it was transmitting the  
sample No. that is selected in  
online menu and it was after  
the designated time.  
It is ended the  
session of transfer  
of the text of test  
results.  
It is output the  
alarm and ended  
The session of the  
transfer of the  
text of test results.  
This text is transmitted when  
the communication is canceled  
in the online menu.  
This text is transmitted when  
the communication is canceled  
by the online alarm.  
Note 1) The number of times a TRI Inquiry can be sent when an online error is generated can be selected in the  
[PARAMETER] [ONLINE] [PROTOCOL] menu.  
- 41 -  
(5.3) Others  
(1) Mixing the session of the TRT inquiry and the session of the test results.  
A. General sequence  
Rx  
Dx  
Dx  
Rx  
AU2700/5400  
Host  
・・・  
Sx  
(a)  
Sx  
Sx  
Sx  
(b)  
(2) Detail  
Section  
(a)  
Definition of section  
Limitation  
The start of the transmission of the text  
of T.R.I. inquiry  
Don't transmit the Test result text.  
-- The end of the receiving of all T.R.I.  
for one samples  
(b)  
The start of the transmission of the text  
of the test results  
Don't transmit the T.R.I. text.  
-- The end of the receiving of all result  
for one samples  
- 42 -  
6. Specification of Connection  
(6.1) In/Output signals and Terminals  
PC Terminal  
at AU2700/5400  
Terminal  
at host  
Signal Name  
Abbrv.  
Direction  
computer  
(25 Pin Configuration)  
-------------------------------------------------------------------------------------------------------------------  
Ground  
FG  
-
1
Data sent  
Data receive  
Signal Ground  
TxDATA  
RxDATA  
SG  
3
2
5
3 (Rx DATA)  
2 (Tx DATA)  
7
(SG)  
Request to send RTS  
Clear to send CTS  
7
8
Caution: RTS and CTS must be connected directly.  
(6.2) Signal Level  
Signal/Signal Format  
Signal Level  
------------------------------------------------------------------------------------  
SPACE (ON)  
MARK (OFF)  
+3V and higher ("H")  
-3V and lower ("L")  
Note: "H" stands for "HIGH LEVEL" and "L" for "LOW LEVEL".  
(6.3) Applicable Connector  
Style of connector : Confirm the standard of IBM PC (RS232 DB9 Male)  
Provide a RS232 DB9 Female connector.  
Style of connector Cover : ditto  
(6.4) Others  
A. When powering ON or OFF the host computer system, ensure that the AU2700/5400 is in  
Standy mode. Powering ON or OFF the host computer system while the AU2700/5400 is being operated may  
result in abnormality function of the AU2700/5400.  
B. The cable for host communication is not included in the AU2700/5400 standard accessories.  
A shielded cable must be used for host connection.  
C. Do not use the terminals on the AU2700/5400 other than No. 2,3,5,7, or 8 for any purpose.  
- 43 -  
A. Appendix  
(1) Flags list  
Flag  
Contents  
(
%
?
Failed to evade the contamination since the detergent for evading the contamination is not enough.  
The sample probe is clogged with sample.  
Calculation unable due to abnormal photometric data  
#
R
!
Sample level detection error  
Reagent level detection error  
Calculation unable due to abnormal data  
Reagent absorbance value at P-START of Reagent Blank run, is smaller than the  
Lower limit of the Parameter.  
Reagent absorbance value at P-START is lower than the lower limit specified  
In the Parameters in routine run.  
U
u
Reagent absorbance value at P-END of Reagent Blank run, is greater than the  
Upper limit of the Parameter.  
Y
Reagent absorbance value at P-END is higher than the upper limit specified  
In the Parameters in routine run.  
y
@
$
Abnormally high result; absorbance of every wavelength is more than 2.5.  
No linearity validation conducted because less than 3 data obtained in the  
Kinetics.  
D
B
Too quick reaction slope in increasing kinetics, absorbance at P-START is  
Higher than MAX. OD in increasing FIXED assay, or too slow reaction slope  
In decreasing kinetics (= no reaction observed)  
Too quick reaction slope in increasing kinetics, or absorbance at P-END is  
Lower than MIN. OD in increasing FIXED assay.  
Linearity error in kinetics.  
*
Linearity error in kinetics.  
&
Z
F
G
x
Prozone data error (prozone check point data error)  
Prozone error  
Result higher than the dynamic range specified in the Parameters  
Result lower than the dynamic range specified in the Parameters  
Other QC data error  
1
2
QC result beyond the QC value range specified in the Parameters  
Multi-rule QC data error  
3
Multi-rule QC data error  
4
Multi-rule QC data error  
5
Multi-rule QC data error  
6
Multi-rule QC data error  
7
Multi-rule QC data error  
p
Result beyond the panic value specified in the Parameters  
Abnormality found in the Inter-Item Check  
T
P
N
H
L
J
Result higher than DECIDE RANGE designated in parameters.  
Result lower than DECIDE RANGE designated in parameters.  
Higher than the result value range specified in the Parameters  
Lower than the result value range specified in the Parameters  
Result higher than the repeat run range specified in the Parameters  
Result lower than the repeat run range specified in the Parameters  
Result extracted for repeat run  
K
S
- 44 -  
/
Test not performed: test has been requisitioned but not performed due  
To any reason.  
r
transferred data to host  
e
c
a
b
)
edited data  
correct data by manual correction  
On-board stability over  
Calibration stability over or Calibration does not exist.  
Current reagent lot number was different from reagent lot No. when ACAL analysis  
- 45 -  
(2) Online parameters list  
Transmission information  
Contents  
menu  
T.R.I. receive  
Normal sample  
Emergency sample  
Stat sample  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
[PARAMETER] [ONLINE] [SETUP]  
menu.  
Normal Repeat sample  
Emergency Repeat sample  
Stat Repeat sample  
Test result transfer  
Normal sample  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/none  
Emergency sample  
Stat sample  
Stat Fast sample  
Normal Repeat sample  
Emergency Repeat sample  
Stat Repeat sample  
QC sample  
Calibration sample  
Reagent blank sample  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Real-time/batch/none  
Protocol 2  
Contents  
menu  
Error control  
Receive  
Continue/stop  
Continue/stop  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
Transfer  
Protocol –1  
Contents  
menu  
Data format  
character length  
parity bit  
7/8  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
ODD/EVEN/none  
1/2  
stop bit  
Basic text format  
start code 1  
start code 2  
end code 1  
end code 2  
Max text length  
unit no.  
00H-FFH  
00H-FFH  
00H-FFH  
00H-FFH  
256/512/1024  
00-99 space  
use/none  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
use ETB  
Sequence control  
bit/sec  
protocol  
BCC check  
retry times  
time out T1  
T2  
4800/9600  
class A/class B  
use/none  
(0-3)  
nn  
nn  
T3  
nn  
T4  
nn  
T5  
nn  
T6  
nn  
T7  
nn  
Text Format  
- 46 -  
Text information  
Contents  
menu  
rack no. /cup position  
rack no. digit  
use/none  
4/5  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
Sex  
age/month  
use/none  
use/none  
[PARAMETER] [FORMAT]  
[REQUISITION FORMAT]  
menu.  
patient information 1  
patient information 2  
patient information 3  
patient information 4  
patient information 5  
patient information 6  
sample ID digits  
use/none digits  
use/none digits  
use/none digits  
use/none digits  
use/none digits  
use/none digits  
4-26 digits  
[PARAMETER] [FORMAT]  
[REQUISITION FORMAT]  
menu.  
[PARAMETER] [FORMAT]  
[REQUISITION FORMAT]  
menu.  
data format  
6 /9 digits  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
data zero suppress  
Sample Type  
used/none  
[PARAMETER] [ONLINE]  
[PROTOCOL] menu.  
used/none  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
[PARAMETER] [ONLINE]  
[PROTOCOL] [SPECIAL] menu.  
Dummy  
(Test Result Transmission Text)  
Block Identification No.  
used/none  
used/none  
ANL Unit No.  
Cuvette Side  
used/none Note 3  
used/none Note 3  
S.No. / S.ID  
S.No. / S.ID  
used/none  
Sequence of S.No. and S.ID in  
T.R.I text  
Sequence of S.No. and S.ID in  
Test result text  
Lot No. and Bottle No.  
Note 1)When you send the “Block Identification No.” field as none, the program will change the parameters automatically as  
following.  
End Code = ETX  
ETB Control = used  
Note 3) Only a Normal sample and repeat sample can be changed.  
- 47 -  
Each of the following parameters are influenced as shown below.  
RB R_ RH RE S_ SH SE DB D_ d_ DH dH DR DA DQ DE  
Sample Type  
Dummy  
-
-
-
-
-
-
-
-
-
-
-
×
×
×
×
×
×
×
×
×
×
×
×
×
×
(Test Result  
Transmission Text)  
Block Identification  
No.  
Sequence of S.No.  
and S.ID in  
T.R.I text  
-
-
-
-
-
-
-
-
-
-
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
-
-
-
-
Sequence of S.No.  
and S.ID in  
Test result text  
ANL Unit No.  
Cuvette Side  
Lot No. and Bottle  
No.  
-
-
-
-
-
-
-
-
-
×
×
×
×
×
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
×
×
×
×
×
×
×
-
: influenced  
: not influenced  
: not apply  
- 48 -  
(3) Character Table  
HIGH  
LOW  
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
0
1
2
3
4
5
6
7
8
9
A
B
DE SP  
0
@
P
p
q
r
s
t
u
v
w
x
y
z
{
SH D1  
SX D2  
EX D3  
ET D4  
EQ NK  
AK SN  
BL EB  
BS CN  
!
"
1
2
3
4
5
6
7
8
9
:
A
B
C
D
E
F
G
H
I
Q
R
S
a
b
c
d
e
f
g
h
i
#
$
%
&
'
T
U
V
W
X
Y
Z
[
(
HT EM )  
LF SB  
*
J
K
j
k
H
EC  
+
;
M
C
D
E
F
CL FS  
CR GS  
SO RS  
SI US  
,
-
.
<
=
>
?
L
\
]
^
_
l
|
}
~
DL  
M
N
O
m
n
o
/
From 20H to 7EH: These characters can be used for the Sample ID code.  
From 30H to 39H: These characters can be used for Test numbers.  
A1 - DC is Japanese character set.  
- 49 -  
(4) AU2700/5400 Online Parameter Sheet  
Set up  
Setting  
T.R.I. Receive  
Routine Normal  
real time batch none  
real time batch none  
real time batch none  
real time batch none  
real time batch none  
real time batch none  
Routine Emergency  
Routine Repeat  
Emergency Repeat  
Stat Normal  
Stat Repeat  
Results Transfer  
Routine Normal  
real time batch none  
Routine Emergency  
Routine Repeat  
Emergency Repeat  
Stat Normal  
Stat Repeat  
real time batch none  
real time batch none  
real time batch none  
real time batch none  
real time batch none  
real time none  
Stat Fast  
QC  
Calibration  
Reagent blank  
real time batch none  
real time batch none  
real time batch none  
Setting  
Upper protocol  
Error control  
T.R.I. Receive  
Results Transfer  
Lower protocol  
Character format  
Character length.  
continue stop  
continue stop  
Setting  
7 8  
Parity bit  
Stop bit  
Data format  
Start code 1  
Start code 2  
End code 1  
End code 2  
Text length  
Unit No.  
ETB control  
SEQ. Control  
Bit/sec  
odd even none  
1 2  
[ ] (01-1F)  
[ ] (00-1F)  
[ ] (01-1F)  
[ ] (00-1F)  
256 512 1024  
[ ] (00-99 space)  
yes no  
4800 9600  
Class A Class B  
yes no  
[ ] (0-3)  
Class  
BCC check  
Retry  
Time out T1  
T2  
[ ] (0-99)  
[ ] (0-99)  
[ ] (0-99)  
[ ] (0-99)  
[ ] (0-99)  
[ ] (0-99)  
[ ] (0-99)  
Setting  
T3  
T4  
T5  
T6  
T7  
Text format  
- 50 -  
Rack No.  
Sex  
none 4 5  
yes no  
AGE/MONTH  
patient inf. 1  
patient inf. 2  
patient inf. 3  
patient inf. 4  
patient inf. 5  
patient inf. 6  
ID digits  
yes no  
yes[ ]digits no  
yes[ ]digits no  
yes[ ]digits no  
yes[ ]digits no  
yes[ ]digits no  
yes[ ]digits no  
[ ]digits (4-26)  
Data format  
ZERO Suppress  
Sample Type  
Dummy  
Block Identification  
ANL unit No.  
Cuvette side  
6 9  
yes no  
yes no  
yes no  
yes no  
yes no  
yes no  
Sequence of S.No. and S.ID in No. S.ID  
T.R.I text  
Sequence of S.No. and S.ID in yes no  
Test result text  
Lot No. and Bottle No.  
yes no  
- 51 -  
(5) ONLINE ALARM LIST  
1)ONLINE ERROR [ aa ] [ bbbbbbbbb ]  
<Operation>  
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated  
as "STOP", then the following will occur:  
1) Text following the alarm will not be received.  
2) In a case when the text requisition is received real-time,the "STOP" status is ignored at the start of  
the next analysis and the text requisition will be received real-time again.  
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated to "CONTINUE",  
the test requisition for the next sample will be received real-time.  
<Description>  
(1).Communication error occurs when transmittng the text for test requisition or when receiving the test requisition format.  
(2).The display in the [ ] and its meaning are as follows.  
aa : type of the errors  
----------------------------------------------------------------  
01 : Device name error  
02 : Framing error  
03 : Over run error  
04 : Parity error  
05 : Time out error  
06 : Receive NAK when transmitting a text  
07 : BCC error when receiving a text  
08 : Other communication errors  
09 : Function error (Full Duplex instead of Half Duplex)  
10 : Unit name error  
11 : Parameter error  
12 : Request cancel  
----------------------------------------------------------------  
bbbbbbbbb  
: Text having error  
----------------------------------------------------------------  
RB  
: Start text for receiving the test  
requisition information.  
RE  
: End text for receiving the test  
requisition information.  
R_ __ nnnn  
: normal run, normal sample, sample No.  
R_ _E nnnn  
R_ _P nnnn  
R_ U_ nnnn  
R_ UE nnnn  
R_ UP nnnn  
R_ X_ nnnn  
R_ XE nnnn  
R_ XP nnnn  
RH __ nnnn  
RH _E nnnn  
RH _P nnnn  
RH U_ nnnn  
RH UE nnnn  
RH UP nnnn  
: normal run, emergency sample, sample No.  
: normal run, STAT sample, sample No.  
: normal run, urine sample, sample No.  
: normal run, urine emergency sample, sample No.  
: normal run, urine STAT sample, sample No.  
: normal run, urine sample, sample No.  
: normal run, urine emergency sample, sample No.  
: normal run, urine STAT sample, sample No.  
: repeat run, normal sample, sample No.  
: repeat run, emergency sample, sample No.  
: repeat run, STAT sample, sample No.  
: repeat run, urine sample, sample No.  
: repeat run, urine emergency sample, sample No.  
: repeat run, urine STAT sample, sample No.  
R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID  
R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID  
R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID  
R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID  
R_ UE xxxxxxxxxxx: normal run, urine emergency sample,sample ID  
R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID  
R_ X_ xxxxxxxxxxx: normal run, urine sample, sample ID  
R_ XE xxxxxxxxxxx: normal run, urine emergency sample,sample ID  
R_ XP xxxxxxxxxxx: normal run, urine STAT sample, sample ID  
RH __ xxxxxxxxxxx: repeat run, emergency sample, sample ID  
RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID  
RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID  
- 52 -  
RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID  
RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID  
RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID  
RH X_ xxxxxxxxxxx: repeat run, emergency sample, sample ID  
RH XE xxxxxxxxxxx: repeat run, emergency sample, sample ID  
RH XP xxxxxxxxxxx: repeat run, STAT sample, sample ID  
----------------------------------------------------------------  
- 53 -  
2)ONLINE ERROR [ aa ] [ bbbbbbbbb ]  
<Operation>  
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",  
the following events can occur:  
1). The text following the alarm will not be received.  
2). In a case when the text requisition is received real-time,the "STOP" designation is ignored and the text requisition  
will be received in real-time at the start of the next analysis.  
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",  
the test requisition for the next sample will be received.  
<Description>  
(1). Communication error occurs when transmitting the text for test requisition or receiving the test requisition format.  
(2). The display in the [ ] and its meaning are as follows.  
aa : type of the errors  
----------------------------------------------------------------  
01 : Device name error  
02 : Framing error  
03 : Over run error  
04 : Parity error  
05 : Time out error  
06 : Receive NAK when transmitting a text  
07 : BCC error when receiving a text  
08 : Other communication errors  
09 : Function error (Full Duplex instead of Half Duplex)  
10 : Unit name error  
11 : Parameter error  
12 : Request cancel  
----------------------------------------------------------------  
bbbbbbbbb  
: Text having error  
----------------------------------------------------------------  
RB  
RE  
: Start text for receiving the test  
requisition information.  
: End text for receiving the test  
requisition information.  
R_ __ nnnn  
R_ _E nnnn  
R_ _P nnnn  
R_ U_ nnnn  
R_ UE nnnn  
R_ UP nnnn  
R_ X_ nnnn  
R_ XE nnnn  
R_ XP nnnn  
RH __ nnnn  
RH _E nnnn  
RH _P nnnn  
: normal run, normal sample, sample No.  
: normal run, emergency sample, sample No.  
: normal run, STAT sample, sample No.  
: normal run, urine sample, sample No.  
: normal run, urine emergency sample, sample No.  
: normal run, urine STAT sample, sample No.  
: normal run, Other normal sample, sample No.  
: normal run, Other emergency sample, sample No.  
: normal run, Other STAT sample, sample No.  
: repeat run, normal sample, sample No.  
: repeat run, emergency sample, sample No.  
: repeat run, STAT sample, sample No.  
RH U_ nnnn  
RH UE nnnn  
RH UP nnnn  
RH X_ nnnn  
RH XE nnnn  
RH XP nnnn  
: repeat run, urine sample, sample No.  
: repeat run, urine emergency sample, sample No.  
: repeat run, urine STAT sample, sample No.  
: repeat run, Other normal sample, sample No.  
: repeat run, Other emergency sample, sample No.  
: repeat run, Other STAT sample, sample No.  
R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID  
R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID  
R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID  
R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID  
R_ UE xxxxxxxxxxx: normal run, urine emergency sample, sample ID  
R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID  
R_ X_ xxxxxxxxxxx: normal run, Other normal sample, sample ID  
R_ XE xxxxxxxxxxx: normal run, Other emergency sample, sample ID  
R_ XP xxxxxxxxxxx: normal run, Other STAT sample, sample ID  
RH __ xxxxxxxxxxx: repeat run, normal sample, sample ID  
RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID  
- 54 -  
RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID  
RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID  
RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID  
RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID  
RH X_ xxxxxxxxxxx: repeat run, Other normal sample, sample ID  
RH XE xxxxxxxxxxx: repeat run, Other emergency sample, sample ID  
RH XP xxxxxxxxxxx: repeat run, Other STAT sample, sample ID  
----------------------------------------------------------------  
- 55 -  
3)ONLINE FORMAT ERROR  
<<Operation>>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for samples following the alarm.  
<<Description>>  
(1).In the Online Parameter menu, submenu Set Up, the T.R.I. receive is not designated asn "YES" for the necessary parameter.  
(2).The display in the [ ] and its meaning are as follows.  
aaaa  
: Type of error  
----------------------------------------------------------------  
R_ __  
R_ _E  
R_ _P  
R_ U_  
R_ UE  
R_ UP  
R_ X_  
R_ XE  
R_ XP  
: normal run, a normal sample  
: normal run, an emergency sample  
: normal run, a STAT sample  
: normal run, a urine normal sample  
: normal run, a urine emergency sample  
: normal run, a urine STAT sample  
: normal run, a Other normal sample  
: normal run, an Other emergency sample  
: normal run, a Other STAT sample  
----------------------------------------------------------------  
RH __  
RH _E  
RH _P  
RH U_  
RH UE  
RH UP  
RH X_  
RH XE  
RH XP  
: repeat run, a normal sample  
: repeat run, an emergency sample  
: repeat run, a STAT sample  
: repeat run, a urine normal sample  
: repeat run, a urine emergency sample  
: repeat run, a urine STAT sample  
: repeat run, a Other normal sample  
: repeat run, an Other emergency sample  
: repeat run, a Other STAT sample  
----------------------------------------------------------------  
bbbbbbbbbb : Information type that is not designated  
No sample No.  
No ID No.  
- 56 -  
4)ONLINE ILLEGAL TEXT CODE [ aa ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for samples following the alarm.  
<Description>  
(1).When receiving the test requisition information text using online, the text classification code is out of specification.  
(2).The display of [ ] and its meaning are as follows.  
aa  
:
----------------------------------------------------------------  
R_  
: Normal sample  
RH  
: Repeat run sample  
----------------------------------------------------------------  
5)ONLINE ILLEGAL TEXT BLOCK No. [ aa -> bb ]  
<Operation>  
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).System will not receive text for samples following the sample generating the alarm.  
3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text  
will be received with real-time again during the next analysis.  
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).The test requisition information text for the next sample will be received.  
<Description>  
(1).When receiving the requisition information text using online, the block identification No. is out of specification.  
(2).The display of [ ] and its meaning are as follows.  
aa : Previously received block identification No.  
bb : Last received block identification No.  
- 57 -  
6)ONLINE ILLEGAL SAMPLE No. [ aa: bbbbb ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for other samples within the run.  
<Description>  
(1).When receiving the test requisition information text using online, sample No. of the sample generating the alarm  
does not meet the specification.  
(2).The display of [ ] and its meaning are as follows.  
aa  
: Sample type  
----------------------------------------------------------------  
__  
: Normal sample  
_E  
_P  
: Emergency sample  
: Stat sample  
U_  
UE  
UP  
X_  
XE  
XP  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other Normal sample  
: Other Emergency sample  
: Other Stat sample  
----------------------------------------------------------------  
bbbb : Sample No.  
----------------------------------------------------------------  
0001-9999 : Normal sample  
E001-E999 : Emergency  
P001-P999 : Stat sample  
----------------------------------------------------------------  
- 58 -  
7)ONLINE ILLEGAL RACK No. [ aaaaaaa: b: cccccccc ]  
<Operation>  
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).System will not receive text for samples following the sample generating the alarm.  
3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text  
will be received with real-time again during the next analysis.  
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).The test requisition information text for the next sample will be received.  
<Description>  
(1).When receiving the requisition information text using online, the rack No. is out of specification.  
(2).The display of [ ] and its meaning are as follows.  
aaa  
: Transferred sample type and sample No.  
---------------------------------------------------------------  
__ nnnn  
_E nnnn  
_P nnnn  
U_ nnnn  
UE nnnn  
UP nnnn  
X_ nnnn  
XE nnnn  
XP nnnn  
bbb  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
: Received sample type and sample No.  
---------------------------------------------------------------  
__ nnnn  
_E nnnn  
_P nnnn  
U_ nnnn  
UE nnnn  
UP nnnn  
X_ nnnn  
XE nnnn  
XP nnnn  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
---------------------------------------------------------------  
: Received the kind of rack  
----------------------------------------------------------------  
b
N
E
R
: Normal sample rack  
: Emergency rack  
: Repeat run rack  
----------------------------------------------------------------  
ccccccc : Received rack No. and position in rack  
- 59 -  
8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for the other samples within the run.  
<Operation>  
(1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm  
does not meet the specification.  
(2).The display of [ ] and its meaning are as follows.  
aa  
: Sample type  
----------------------------------------------------------------  
__  
: Normal sample  
_E  
_P  
: Emergency sample  
: Stat sample  
U_  
UE  
UP  
X_  
XE  
XP  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other stat sample  
----------------------------------------------------------------  
9)ONLINE ILLEGAL GENDER TEXT [ a ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for other samples within the run.  
<Description>  
(1).When receiving the requisition information text using online, the sex of this text is not within the specification.  
(2).The display of [ ] and its meaning are as follows.  
a
: Received sex  
10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for other samples within the run.  
<Description>  
(1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this  
text is out of the specification.  
(2).The display of [ ] and its meaning are as follows.  
aaa  
bb  
: Received age  
: Received month age  
- 60 -  
11)ONLINE ANALYSIS METHOD MISMATCH [ aa <> bb ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for other samples within the run.  
<Description>  
(1).When receiving the test requisition information text using online, sample type of this text does not match  
the required sample type.  
(2).The display of [ ] and its meaning are as follows.  
aa  
: Sample type  
----------------------------------------------------------------  
__  
: Normal sample  
_E  
_P  
: Emergency sample  
: Stat sample  
U_  
UE  
UP  
X_  
XE  
XP  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
----------------------------------------------------------------  
bb : Received sample type  
----------------------------------------------------------------  
__  
: Normal sample  
_E  
_P  
: Emergency sample  
: Stat sample  
U_  
UE  
UP  
X_  
XE  
XP  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
----------------------------------------------------------------  
12)ONLINE SAMPLE No. MISMATCH [ aaaaaa <> bbbbbb ]  
- 61 -  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for the other samples within the run.  
<Description>  
(1).When receiving the test requisition information text using online, sample type of this does not match the required sample.  
(2).The display of [ ] and its meaning are as follows.  
aaaaaaa  
: Transferred sample type and sample No.  
---------------------------------------------------------------  
__ nnnn  
_E nnnn  
_P nnnn  
U_ nnnn  
UE nnnn  
UP nnnn  
X_ nnnn  
XE nnnn  
XP nnnn  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
---------------------------------------------------------------  
bbbbbb : Received sample type and sample No.  
---------------------------------------------------------------  
__ nnnn  
_E nnnn  
_P nnnn  
U_ nnnn  
UE nnnn  
UP nnnn  
X_ nnnn  
XE nnnn  
XP nnnn  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
---------------------------------------------------------------  
- 62 -  
13)ONLINE RACK No. MISMATCH [ aaaaaa : bbbb <> cccc ]  
<Operation>  
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).System will not receive text for samples following the sample generating the alarm.  
3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text  
will be received with real-time again during the next analysis.  
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",  
the following events occur:  
1).Disregard the test requisition information text for the sample generating the alarm.  
2).The test requisition information text for the next sample will be received.  
<Description>  
(1).When receiving the test requisition information text using online, rack No. of this does not match the required sample.  
(2).The display of [ ] and its meaning are as follows.  
aaaaaaaaa : Transferred sample type and sample No.  
---------------------------------------------------------------  
__ nnnn  
_E nnnn  
_P nnnn  
U_ nnnn  
UE nnnn  
UP nnnn  
X_ nnnn  
XE nnnn  
XP nnnn  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
---------------------------------------------------------------  
bbbb : Transferred rack No.  
---------------------------------------------------------------  
cccc : Received rack No.  
---------------------------------------------------------------  
14)ONLINE MISMATCH [ aaaaaaaaaa <> bbbbbbbbbbb ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for the other samples within the run.  
<Description>  
(1).When receiving the test requisition information text using online, sample ID of this text does not match  
the required ones.  
(2).The display of [ ] and its meaning as follows.  
aaaaaaaaa  
bbbbbbbbb  
: Transferred sample ID  
: Received sample ID  
- 63 -  
15)ONLINE ITEM ERROR [ aaaaaaaa ]  
<Operation>  
(1).Disregard the test requisition information text for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for the other samples within the run.  
<Description>  
(1).When receiving the test requisition information text using online, the test requisition cannot be stored.  
(2).The display of [ ] and its meaning are as follows.  
aaaaaa  
: Sample type and sample No.  
---------------------------------------------------------------  
__nnnn  
: Normal sample  
__Ennn  
__Pnnn  
_U_nnn  
_UEnnn  
_UPnnn  
_Xnnnn  
_XEnnn  
_XPnnn  
H_nnnn  
H_Ennn  
H_Pnnn  
HUnnnn  
HUEnnn  
HUPnnn  
HXnnnn  
HXEnnn  
HXPnnn  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
: repeat run normal sample  
: repeat run emergency sample  
: repeat run stat sample  
: repeat run urine normal sample  
: repeat run urine emergency sample  
: repeat run urine stat sample  
: repeat run other normal sample  
: repeat run other emergency sample  
: repeat run other stat sample  
---------------------------------------------------------------  
- 64 -  
16)ONLINE RERUN ITEM ERROR [ aaaaaaaa ]  
<Operation>  
(1).Disregard the repeat run test requisition information for the sample generating the alarm.  
(2).Continue to receive the test requisition information text for the other samples within the run.  
<Description>  
(1).When receiving the test requisition information text for repeat run using online, repeat run test requisition cannot be  
stored with for the following reasons.  
1).Original sample that is designated in the repeat run text is not stored in the normal run/sample test requisition  
information.  
2).Sample No. is not set to the original sample that is designated in the repeat run text.  
3).Original sample that is designated in the repeat run text has already been stored as a repeat run sample  
for another sample.  
(2).The display of [ ] and its meaning are as follows.  
----------------------------------------------------------------  
aaaaaa  
: sample type and original sample No.  
: Normal sample  
: Emergency sample  
: Stat sample  
: Urine normal sample  
: Urine emergency sample  
: Urine stat sample  
: Other normal sample  
: Other emergency sample  
: Other Stat sample  
__nnnn  
__Ennn  
__Pnnn  
_Unnnn  
_UEnnn  
_UPnnn  
_Xnnnn  
_XEnnn  
_XPnnn  
----------------------------------------------------------------  
- 65 -  

MTD Shredder 770 95 7B User Manual
Motorola I296 H88XAH6JR2AN User Manual
Lindy 42791 User Manual
LG Electronics VENUS User Manual
Dell Latitude Computer L400 User Manual
Capresso AD10BLK User Manual
Bunn Single 270387000H User Manual
Bookendz BE MBP17 User Manual
Aluratek ACSSDM64F User Manual
Addonics Technologies ZHDUESA User Manual