ATD Tools Clock ATD 701 User Manual

Warning  
Do not point laser at eye or indirectly off reflective surfaces.  
1. When take measurement, point thermometer toward  
the object to be measured and hold the red trigger. The  
object under test should be larger than the spot size  
calculated by the field of view diagram.  
ATD-701  
Non-contact Infrared Thermometer  
Instruction manual  
2. Distance spot size: As the distance from the object  
increase, the spot size of measuring area becomes  
larger.  
3. Field of view: Make sure the target is larger than the  
unit’s spot size. The smaller the target, the closer  
measuring distance. When accuracy is critical, make  
sure the target is at least twice as large as the spot size.  
4. Emissivity: Most organic materials and painted or  
oxidized surfaces have an emissivity of 0.95. Inaccurate  
readings will result from measuring shiny or polished  
metal surfaces. To compensate, cover the surface to be  
measured with masking tape or flat black paint.  
Measure the tape or painted surface when the tape or  
painted reach the same temperature as the material  
underneath.  
1. Introduction  
Compact, rugged and easy to use. Just aim and push the  
button, read target surface temperatures in less than a  
second. Safely measure surface temperatures of hot,  
hazardous or hard-to-reach objects without contact.  
2. Quick start instruction  
How it works  
Infrared thermometer measures the surface temperature of  
an object. The unit’s optics sense emitted, reflected, and  
transmitted energy which is collect and focused onto a  
detector. The unit’s electronics transmitted energy which is  
display on the unit. For increased ease and accuracy the  
laser pointer makes aiming even more precise.  
Cautions  
1. Press battery door clip, install battery correctly. Pull the  
trigger, LCD display reading battery icon. Release  
the trigger and the reading will hold for 7 seconds.  
LCD display:  
Infrared thermometer should be protected from the  
following:  
--EMF (electro-magnetic fields) from arc welders, induction  
heaters.  
--Thermal shock (cause by large or abrupt ambient  
temperature changes allow 1 hour for unit to stabilize before  
use).  
--Do not leave the unit on or near objects of high  
temperature.  
A
B
C
D
E
F
Laser pointer turn on prompt  
Alarm of high temperature prompt  
Alarm of Low temperature prompt  
Data Hold  
Measure result  
Emissivity  
G
H
Temperature unit:℃  
Temperature unit:℉  

Motorola WT4090 User Manual
Marantz PMD340 User Manual
Kenwood DPF J5030 User Manual
ID Tech 25 User Manual
Hitachi Travelstar HTS725025A7E635 User Manual
GPX C13807 User Manual
Bodum MOCCA 1176 USA User Manual
AT&T MOTO Z9 User Manual
Antec Quiet Computing 2U26EPS600 User Manual
Alpine CDA 9827 User Manual